CROSSCHECK(R) Plating system, HOCKEY STICK CX Plate, Left, REF CCP-HSX1L. Orthopedic use.
FDA recall event 70859 · Class II · Wright Medical Technology, Inc.
Key facts
| Event ID | 70859 |
|---|---|
| Recall numbers | Z-1482-2015, Z-1483-2015, Z-1484-2015, Z-1485-2015, Z-1486-2015, Z-1487-2015, Z-1488-2015, Z-1489-2015, Z-1490-2015, Z-1491-2015, Z-1492-2015, Z-1493-2015 and 7 more |
| Recalling firm | Wright Medical Technology, Inc. |
| Product type | Devices |
| Products | CROSSCHECK(R) Plating system, HOCKEY STICK CX Plate, Left, REF CCP-HSX1L. Orthopedic use., CROSSCHECK(R) Plating system, HOCKEY STICK CX Plate, Right, REF CCP-HSX1R. Orthopedic use., CROSSCHECK(R) Plating system, LAPIDUS CX Plate, 1mm Step, Left, REF CCP-LPX1L. Orthopedic use., CROSSCHECK(R) Plating system, LAPIDUS CX Plate, 1mm Step, Right, REF CCP-LPX1R. Orthopedic use., CROSSCHECK(R) Plating system, LAPIDUS CX Plate, 2mm Step, Left, REF CCP-LPX2L. Orthopedic use. and 14 more |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction, The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction. |
| Initiation date | Mar 6, 2015 |
| Report date | Apr 29, 2015 |
| Termination date | Aug 5, 2018 |
| City | Memphis |
| State | TN |
| Country | United States |
| FEI number | 3010667733 |
At a glance
- 10–99 Products in this recall
- 1–9 FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About FDA drug and product recalls
FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.
Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: