Transonic Hemodialysis Monitor, Model HD03 -- Product Usage: The Hemodialysis Monitor is intended for use by…

FDA recall event 70647 · Class II · Transonic Systems Inc

Key facts

Event ID 70647
Recall numbers Z-1295-2015
Recalling firm Transonic Systems Inc
Product type Devices
Products Transonic Hemodialysis Monitor, Model HD03 -- Product Usage: The Hemodialysis Monitor is intended for use by trained medical personnel to measure delivered blood flow, vascular access recirculation…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Customer unable to obtain measurements due to the device seeing no flow or irrelevant flow condition during dialysis procedures. The user saw a zero or a very low flow value displayed on HD03 and/or…
Initiation date Feb 17, 2015
Report date Mar 25, 2015
Termination date May 31, 2016