Drug recall
GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number GAP-N48-16; GAP-N48-18…
FDA recall event 70630 · Class II · Pega Medical Inc.
Key facts
| Event ID | 70630 |
|---|---|
| Recall numbers | Z-1362-2015 |
| Recalling firm | Pega Medical Inc. |
| Product type | Devices |
| Products | GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number GAP-N48-16; GAP-N48-18; GAP-N48-26; GAP-N48-28. Product Usage: The GAP Endo-Exo Medullary System is used for the… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Failures of the 4.8mm of the GAP Endo-Exo Medullary System have been identified at the level of the distal cortical screw holes. Failures mainly occurred in patients with tibial pseudoarthrosis (CPT)… |
| Initiation date | Feb 20, 2015 |
| Report date | Apr 8, 2015 |
| Termination date | Apr 30, 2015 |
| City | Laval |
| Country | Canada |
| FEI number | 3000327445 |