Drug recall

Telluride Percutaneous Rod Inserter Long, Model #7706-1904. The Inserter is a reusable device provided within…

FDA recall event 70376 · Class II · Biomet Spine, LLC

Key facts

Event ID 70376
Recall numbers Z-1089-2015
Recalling firm Biomet Spine, LLC
Product type Devices
Products Telluride Percutaneous Rod Inserter Long, Model #7706-1904. The Inserter is a reusable device provided within an aluminum sterilization case along with other orthopedic manual surgical instruments…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The Telluride Percutaneous Rod Inserter Long does not contain a feature that locks the Rod in place during use. There is a potential for the Rod to disengage from the Inserter during insertion, and…
Initiation date Dec 19, 2011
Report date Feb 18, 2015
Termination date Mar 23, 2015
City Broomfield
State CO
Country United States
FEI number 3004485144

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls