Drug recall

The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the…

FDA recall event 70271 · Class II · Medtronic Neuromodulation

Key facts

Event ID 70271
Recall numbers Z-1147-2015
Recalling firm Medtronic Neuromodulation
Product type Devices
Products The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or muscles when…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Current labeling indicates that cycling improves device longevity in every programming scenario. However, enabling cycling at certain parameter settings may decrease the device battery longevity. With…
Initiation date Jan 9, 2015
Report date Mar 4, 2015
Termination date Sep 28, 2017
City Minneapolis
State MN
Country United States
FEI number 3007566237

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)