Drug recall

The Arrow Edge Hemodialysis Catheterization Set Long-Term Access and NextStep Antegrade Hemodialysis…

FDA recall event 69938 · Class II · Arrow International Inc

Key facts

Event ID 69938
Recall numbers Z-1007-2015
Recalling firm Arrow International Inc
Product type Devices
Products The Arrow Edge Hemodialysis Catheterization Set Long-Term Access and NextStep Antegrade Hemodialysis Catheterization Set, Long-Term Access, part numbers: AC-15192-SFX, AC-15232-SFX, AC-15272-SFX…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The tunneling sheath packaged within the kits can crack during placement. If the tunneling sheath cracks, it can cause a delay in treatment for the time it takes to retrieve a replacement kit.
Initiation date Feb 4, 2013
Report date Jan 28, 2015
Termination date Oct 14, 2016
City Reading
State PA
Country United States
FEI number 2518433

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: