Drug recall

Philips Multi Diagnost Eleva II

FDA recall event 69870 · Class II · Philips Medical Systems, Inc.

Key facts

Event ID 69870
Recall numbers Z-1044-2015, Z-1045-2015
Recalling firm Philips Medical Systems, Inc.
Product type Devices
Products Philips Multi Diagnost Eleva II, Philips Multi Diagnost Eleva with FlatDetector
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The possibility of the Reference Air Kerma Rate for low dose mode will not be reduced to 50% of the value for the normal mode, but will be approximately 70-80%, leading to a higher patient dose than…
Initiation date Oct 14, 2013
Report date Mar 4, 2015
Termination date Jan 23, 2017
City Andover
State MA
Country United States
FEI number 1218950

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)