DPM 6 Patient Monitor and DPM 7 Patient Monitor. The product affected is the Mindray IBP module, P/N is…

FDA recall event 69788 · Class II · Mindray DS USA, Inc. dba Mindray North America

Key facts

Event ID 69788
Recall numbers Z-0837-2015
Recalling firm Mindray DS USA, Inc. dba Mindray North America
Product type Devices
Products DPM 6 Patient Monitor and DPM 7 Patient Monitor. The product affected is the Mindray IBP module, P/N is 6800-30-50485 for use with Mindray DPM 6 and DPM 7 patient monitors . Intended to be used for…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Mindray has identified an issue that involves the IBP module used with the DPM 6 and DPM 7 Monitor where the IBP measurement data may be displayed in an incorrect IBP tile on the monitors display.
Initiation date Oct 6, 2014
Report date Dec 31, 2014
Termination date Jun 6, 2017