Drug recall

Breas Vivo 50 Home Care Ventilator Product Usage: Respiratory Care

FDA recall event 69760 · Class II · Breas Medical AB

Key facts

Event ID 69760
Recall numbers Z-2094-2015
Recalling firm Breas Medical AB
Product type Devices
Products Breas Vivo 50 Home Care Ventilator Product Usage: Respiratory Care
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Unintended treatment termination could result from a keypad malfunction in some situations. The device erroneously interprets this as a Stop Treatment Instruction. An alarm will not sound, or be…
Initiation date Nov 7, 2014
Report date Jul 22, 2015
Termination date Apr 7, 2016
City Molnlycke
Country Sweden
FEI number 3003244771

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: