Drug recall

Esprit V1000 Ventilator and Esprit/V200 Conversion Ventilator The Esprit V1000 and Esprit/V200 Conversion…

FDA recall event 69728 · Class II · Respironics California Inc

Key facts

Event ID 69728
Recall numbers Z-0485-2015
Recalling firm Respironics California Inc
Product type Devices
Products Esprit V1000 Ventilator and Esprit/V200 Conversion Ventilator The Esprit V1000 and Esprit/V200 Conversion Ventilators are microprocessor-controlled, electrically powered mechanical ventilators…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the…
Initiation date Oct 29, 2014
Report date Dec 3, 2014
Termination date Jun 7, 2016
City Carlsbad
State CA
Country United States
FEI number 3002470603