Drug recall
GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal…
FDA recall event 69660 · Class II · GE OEC Medical Systems, Inc
Key facts
| Event ID | 69660 |
|---|---|
| Recall numbers | Z-0840-2015, Z-0841-2015 |
| Recalling firm | GE OEC Medical Systems, Inc |
| Product type | Devices |
| Products | GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize., GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize. |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as… |
| Initiation date | Sep 15, 2014 |
| Report date | Jan 21, 2015 |
| Termination date | May 20, 2015 |
| City | Salt Lake City |
| State | UT |
| Country | United States |
| FEI number | 1720753 |