Drug recall

GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal…

FDA recall event 69660 · Class II · GE OEC Medical Systems, Inc

Key facts

Event ID 69660
Recall numbers Z-0840-2015, Z-0841-2015
Recalling firm GE OEC Medical Systems, Inc
Product type Devices
Products GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize., GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as…
Initiation date Sep 15, 2014
Report date Jan 21, 2015
Termination date May 20, 2015
City Salt Lake City
State UT
Country United States
FEI number 1720753

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls