Key facts
| Event ID | 68985 |
|---|---|
| Recall numbers | Z-2581-2014 |
| Recalling firm | Elekta, Inc. |
| Product type | Devices |
| Products | Monaco The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon and electron treatment plans and… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Incorrect Delivery of Composite VMAT Fields - When DICOM is exporting an arc plan (VMAT, Dynamic Conformal or 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported… |
| Initiation date | Aug 18, 2014 |
| Report date | Sep 10, 2014 |
| Termination date | Sep 15, 2021 |