Drug recall

FOCAL Sim Workstation, Radiation Treatment Planning System. Used for patient contouring and viewing of…

FDA recall event 68984 · Class II · Elekta, Inc.

Key facts

Event ID 68984
Recall numbers Z-2574-2014
Recalling firm Elekta, Inc.
Product type Devices
Products FOCAL Sim Workstation, Radiation Treatment Planning System. Used for patient contouring and viewing of treatment plan outputs. Also able to perform CT simulations and forward their information to the…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall When DICOM is exporting an arc plan (VMAT, Dynamic Conformal of 3D Static Arcs) and using the Composite Field Sequencing (CFS) feature, the DICOM exported plan will not match the approved treatment…
Initiation date Aug 18, 2014
Report date Sep 10, 2014
Termination date Aug 8, 2017
City Atlanta
State GA
Country United States
FEI number 1037831

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: