Drug recall

VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000…

FDA recall event 68909 · Class II · MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK

Key facts

Event ID 68909
Recall numbers Z-2653-2014
Recalling firm MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Product type Devices
Products VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become…
Initiation date Jul 5, 2014
Report date Sep 24, 2014
Termination date Apr 1, 2015
City HIROSHIMA
Country Japan
FEI number 3006942329

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls