Drug recall

Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1017. Product Usage…

FDA recall event 68763 · Class II · Medical Positioning Inc

Key facts

Event ID 68763
Recall numbers Z-2208-2014, Z-2209-2014, Z-2210-2014, Z-2211-2014
Recalling firm Medical Positioning Inc
Product type Devices
Products Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1017. Product Usage: Used to conduct syncope examinations., Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1019. Product Usage: Used to conduct syncope examinations., Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1219. Product Usage: Used to conduct syncope examinations., Medical Positioning Rapid Response HUT table with Dual 2 Way Drop Section, Powered Medical Examination table, Model # 1217. Product Usage: Used to conduct syncope examinations.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can…
Initiation date Jul 10, 2014
Report date Aug 20, 2014
Termination date Jan 3, 2017
City Kansas City
State MO
Country United States
FEI number 1000118382

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls