Drug recall

Philips Medical Systems Computed Tomography X-ray systems (Brilliance CT series, iCT, iCT SP, Ingenuity CT…

FDA recall event 68245 · Class II · Philips Medical Systems (Cleveland) Inc

Key facts

Event ID 68245
Recall numbers Z-1722-2014
Recalling firm Philips Medical Systems (Cleveland) Inc
Product type Devices
Products Philips Medical Systems Computed Tomography X-ray systems (Brilliance CT series, iCT, iCT SP, Ingenuity CT, Ingenuity Core, Ingenuity Core128, Ingenuity Flex, and the MX8000 Dual v. EXP scanners)
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.
Initiation date Apr 1, 2014
Report date Jun 11, 2014
Termination date Jun 26, 2018
City Cleveland
State OH
Country United States