Drug recall
RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and…
FDA recall event 68162 · Class II · RAYSEARCH LABORATORIES AB
Key facts
| Event ID | 68162 |
|---|---|
| Recall numbers | Z-1806-2014 |
| Recalling firm | RAYSEARCH LABORATORIES AB |
| Product type | Devices |
| Products | RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy. |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects… |
| Initiation date | Apr 25, 2014 |
| Report date | Jun 25, 2014 |
| Termination date | Jun 26, 2017 |
| City | Stockholm |
| Country | Sweden |
| FEI number | 3007774465 |