Drug recall

Variant II beta-thalassemia Short Program Model Numbers 270-2103 (250 tests) and 270-2154 (500 tests), Bio-Rad…

FDA recall event 68016 · Class II · Bio-Rad Laboratories, Inc.

Key facts

Event ID 68016
Recall numbers Z-1537-2014
Recalling firm Bio-Rad Laboratories, Inc.
Product type Devices
Products Variant II beta-thalassemia Short Program Model Numbers 270-2103 (250 tests) and 270-2154 (500 tests), Bio-Rad Laboratories, Inc. in vitro diagnostic. The Variant II Hemoglobin testing system uses the…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall There is an issue with the buffers which impacts the quantification of the Hemoglobin F peak. Calibration and Quality control are likely to fail if the affected lots are used.
Initiation date Apr 8, 2014
Report date May 14, 2014
Termination date May 30, 2014
City Hercules
State CA
Country United States
FEI number 2915274

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls