Drug recall

Cobas c8000 Modular Analyzer Series Cobas c502, Part Number 5964067001, a clinical chemistry analyzer intended…

FDA recall event 67762 · Class II · Roche Diagnostics Operations, Inc.

Key facts

Event ID 67762
Recall numbers Z-1449-2014
Recalling firm Roche Diagnostics Operations, Inc.
Product type Devices
Products Cobas c8000 Modular Analyzer Series Cobas c502, Part Number 5964067001, a clinical chemistry analyzer intended for the in vitro quantitative / qualitative determination of analytes in body fluids.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The Initial Cassette Volume Check (ICVC) is not activated on the cobas c 502 module of the cobas 8000 modular analyzer series when manually filled cobas c packs are loaded. This issue occurs with…
Initiation date Mar 13, 2014
Report date Apr 16, 2014
Termination date Jan 10, 2018
City Indianapolis
State IN
Country United States
FEI number 1823260

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls