Drug recall

Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116…

FDA recall event 67678 · Class II · Ion Beam Applications S.A.

Key facts

Event ID 67678
Recall numbers Z-1492-2014
Recalling firm Ion Beam Applications S.A.
Product type Devices
Products Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116. Proton therapy
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Internal testing of the Proteus 235 Proton Therapy System revealed incorrect dose rate calculation in case of beam line option other than Zero (0).
Initiation date Dec 4, 2013
Report date Apr 30, 2014
Termination date Jun 12, 2014
City Louvain La Neuve
Country Belgium
FEI number 3000256071