IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872…
FDA recall event 67210 · Class II · Siemens Healthcare Diagnostics
Key facts
| Event ID | 67210 |
|---|---|
| Recall numbers | Z-1517-2014, Z-1518-2014 |
| Recalling firm | Siemens Healthcare Diagnostics |
| Product type | Devices |
| Products | IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diagnostics Products Ltd. For in vitro diagnostic use with the…, IMMULITE /IMMULITE 1000 Systems TIE Total IgE, REF LKIE1, SMN 10380867, IVD --- Siemens Healthcare Diagnostics Products Ltd. For in vitro diagnostic use with the IMMULITE, IMMULITE 1000, and IMMULITE… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays. |
| Initiation date | Dec 20, 2013 |
| Report date | May 7, 2014 |
| Termination date | May 18, 2018 |
| City | Tarrytown |
| State | NY |
| Country | United States |