Drug recall

PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee…

FDA recall event 67159 · Class II · DePuy Orthopaedics, Inc.

Key facts

Event ID 67159
Recall numbers Z-1077-2014
Recalling firm DePuy Orthopaedics, Inc.
Product type Devices
Products PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This…
Initiation date Dec 9, 2013
Report date Mar 5, 2014
Termination date Jun 10, 2014
City Warsaw
State IN
Country United States
FEI number 1818910

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls