Drug recall

GE Healthcare - Giraffe and Panda Bag and Mask Resuscitation System;Giraffe Stand-alone Infant Resuscitation…

FDA recall event 67004 · Class I · GE Healthcare, LLC

Key facts

Event ID 67004
Recall numbers Z-1070-2014, Z-1071-2014, Z-1072-2014
Recalling firm GE Healthcare, LLC
Product type Devices
Products GE Healthcare - Giraffe and Panda Bag and Mask Resuscitation System;Giraffe Stand-alone Infant Resuscitation System. Product Usage: The Bag and Mask Resuscitation System provides the basic equipment…, GE Healthcare - Giraffe¿ and Panda¿ iRes Infant Warmer Product Usage: The infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on…, GE Healthcare - Giraffe Warmer;Panda Warmer;Panda iRes Warmer, Giraffe Stand-alone Infant Resuscitation System. Product Usage: The T-Piece Resuscitation System [ventilator, emergency powered…
Classification Class I
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.
Initiation date Oct 1, 2013
Report date Mar 5, 2014
Termination date Jul 22, 2016
City Waukesha
State WI
Country United States
FEI number 2126677

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls