PedFuse Reset Screw; Model: 01-80115-55 The PedFuse Reset pedicle screw implants intended to be used to help…

FDA recall event 66698 · Class II · SpineFrontier, Inc.

Key facts

Event ID 66698
Recall numbers Z-0544-2014
Recalling firm SpineFrontier, Inc.
Product type Devices
Products PedFuse Reset Screw; Model: 01-80115-55 The PedFuse Reset pedicle screw implants intended to be used to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall During routine servicing, it was discovered that one 8mmx55mm Reset screw was assembled with an 8mm tulip head and a 9mm screw. The implant should have consisted of an 8mm tulip head and an 8mm screw…
Initiation date Oct 18, 2013
Report date Dec 25, 2013
Termination date Aug 22, 2016
City Beverly
State MA
Country United States