Drug recall

The EkoSonic Endovascular System employs high frequency (2-2.5 MHz), low power (0.585 watt average power per…

FDA recall event 66504 · Class II · EKOS Corporation

Key facts

Event ID 66504
Recall numbers Z-0133-2014
Recalling firm EKOS Corporation
Product type Devices
Products The EkoSonic Endovascular System employs high frequency (2-2.5 MHz), low power (0.585 watt average power per transducer) ultrasound to facilitate the delivery of physician-specified fluids in the…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The EkoSonic Control Unit is recalled because it had the incorrect power entry module installed. It is missing an AC filter to reduce the noise and electrical interference.
Initiation date Oct 2, 2013
Report date Nov 13, 2013
Termination date Nov 22, 2013
City Bothell
State WA
Country United States
FEI number 3001627457

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls