Drug recall
Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657…
FDA recall event 66274 · Class II · Leica Microsystems, Inc.
Key facts
| Event ID | 66274 |
|---|---|
| Recall numbers | Z-1725-2014, Z-1726-2014 |
| Recalling firm | Leica Microsystems, Inc. |
| Product type | Devices |
| Products | Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V…, Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6… |
| Initiation date | Sep 11, 2013 |
| Report date | Jun 18, 2014 |
| Termination date | Apr 24, 2017 |
| City | Buffalo Grove |
| State | IL |
| Country | United States |
| FEI number | 3000210133 |
At a glance
- 1–9 Products in this recall
- 1–9 FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About FDA drug and product recalls
FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.