Drug recall

COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692…

FDA recall event 66239 · Class II · Roche Diagnostics Operations, Inc.

Key facts

Event ID 66239
Recall numbers Z-0168-2014
Recalling firm Roche Diagnostics Operations, Inc.
Product type Devices
Products COBAS INTEGRA 800 Analyzer with software 9864.C2 Cobas. COBAS Part Numbers 28122474001, 28122474692, 04559126001, 04589459970 corresponding to the Analyzer with or without closed tube sampling…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Roche COBAS INTEGRA 800 system, Software Version 9864.C2 does not perform necessary ISE Service Actions. Whenever the ISE rack is removed, the ISE counters in the software are reset to zero and the…
Initiation date Sep 16, 2013
Report date Nov 13, 2013
Termination date May 12, 2015
City Indianapolis
State IN
Country United States
FEI number 1823260

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls