Drug recall

DC power cable (p/n 3010556) supplied with the MEDRAD(R) Veris MR Vital Signs Monitor: Product Usage: Monitor…

FDA recall event 65411 · Class II · Medrad Inc

Key facts

Event ID 65411
Recall numbers Z-1586-2013
Recalling firm Medrad Inc
Product type Devices
Products DC power cable (p/n 3010556) supplied with the MEDRAD(R) Veris MR Vital Signs Monitor: Product Usage: Monitor, Physiological, Patient
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The firm is recalling these power cables due to a latent design reliability issue and the potential for shorting which can result in heating/melting of the cable jacket.
Initiation date Apr 15, 2013
Report date Jul 3, 2013
Termination date Apr 3, 2014
City Indianola
State PA
Country United States
FEI number 2520313

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls