Drug recall

Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to be…

FDA recall event 65082 · Class II · Canon Inc.

Key facts

Event ID 65082
Recall numbers Z-1404-2013
Recalling firm Canon Inc.
Product type Devices
Products Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to be used to measure the refractive power and the radius of corneal curvature of the human…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Canon Inc. had received 17 claims from our foreign sales companies between October to November 2012 that measurement value of RK-F2 s refractive power (SPH value) sometimes indicated abnormal value…
Initiation date Dec 5, 2012
Report date Jun 5, 2013
Termination date Sep 20, 2013
City Tokyo
Country Japan
FEI number 1000181430

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls