Drug recall

KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a…

FDA recall event 64989 · Class II · SpineFrontier, Inc.

Key facts

Event ID 64989
Recall numbers Z-1737-2013
Recalling firm SpineFrontier, Inc.
Product type Devices
Products KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles.
Initiation date Dec 30, 2011
Report date Jul 24, 2013
Termination date Jan 8, 2014
City Beverly
State MA
Country United States