Drug recall

cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from…

FDA recall event 64930 · Class II · Roche Molecular Systems, Inc.

Key facts

Event ID 64930
Recall numbers Z-1346-2013
Recalling firm Roche Molecular Systems, Inc.
Product type Devices
Products cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted…
Initiation date Apr 1, 2013
Report date May 29, 2013
Termination date Aug 7, 2014
City Branchburg
State NJ
Country United States
FEI number 2243471

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: