GE, VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspeed VCT, Discovery CT750HD…

FDA recall event 64889 · Class II · GE Healthcare, LLC

Key facts

Event ID 64889
Recall numbers Z-1188-2013
Recalling firm GE Healthcare, LLC
Product type Devices
Products GE, VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspeed VCT, Discovery CT750HD. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly…
Initiation date Jul 25, 2011
Report date May 8, 2013
Termination date Feb 4, 2014
City Waukesha
State WI
Country United States
FEI number 2126677

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: