Drug recall

GemStar Pump - Battery Product Usage: The GemStar Pump is a small and lightweight, single channel infusion…

FDA recall event 64839 · Class I · Hospira Inc.

Key facts

Event ID 64839
Recall numbers Z-1159-2013, Z-1160-2013
Recalling firm Hospira Inc.
Product type Devices
Products GemStar Pump - Battery Product Usage: The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is…
Classification Class I
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Battery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used.
Initiation date Mar 18, 2013
Report date May 8, 2013
Termination date Mar 7, 2017
City Lake Forest
State IL
Country United States
FEI number 3005579246