Drug recall

Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.

FDA recall event 64673 · Class II · Toshiba American Medical Systems Inc

Key facts

Event ID 64673
Recall numbers Z-1068-2013
Recalling firm Toshiba American Medical Systems Inc
Product type Devices
Products Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Due to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error…
Initiation date Nov 14, 2012
Report date Apr 17, 2013
Termination date Apr 8, 2013
City Tustin
State CA
Country United States
FEI number 2020563

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls