Drug recall

GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging systems. For use…

FDA recall event 64274 · Class II · GE Healthcare, LLC

Key facts

Event ID 64274
Recall numbers Z-0934-2013
Recalling firm GE Healthcare, LLC
Product type Devices
Products GE Healthcare, Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ. Cardiovascular X-ray imaging systems. For use in generating fluoroscopic images of human anatomy.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Exposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.
Initiation date Feb 4, 2011
Report date Mar 20, 2013
Termination date Jan 27, 2014
City Waukesha
State WI
Country United States
FEI number 2126677