Drug recall

The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and…

FDA recall event 64263 · Class II · Hospira Inc.

Key facts

Event ID 64263
Recall numbers Z-0283-2014, Z-0284-2014, Z-0285-2014, Z-0286-2014, Z-0287-2014, Z-0288-2014, Z-0289-2014, Z-0290-2014
Recalling firm Hospira Inc.
Product type Devices
Products The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+3…, The Plum A+ Hyperbaric Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The…, The Plum A+ Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal…
Initiation date Feb 1, 2013
Report date Nov 20, 2013
Termination date Dec 6, 2016
City Lake Forest
State IL
Country United States
FEI number 3005579246

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls