Drug recall

Philips Diagnost Stationary Radiographic System Models: 712020, 712022, 712050, 712052, 712055, 712057, 712062…

FDA recall event 64130 · Class II · Philips Healthcare Inc.

Key facts

Event ID 64130
Recall numbers Z-1447-2013, Z-1448-2013
Recalling firm Philips Healthcare Inc.
Product type Devices
Products Philips Diagnost Stationary Radiographic System Models: 712020, 712022, 712050, 712052, 712055, 712057, 712062 Product Usage: Universal RAD Applications
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Installation Acceptance Test Document (IATD) non compliant with regulations IATDs (Installation Acceptance Test could not be located or different templates have been used, Installation Acceptance Test Document (IATD) non compliant with regulations. Records shows table height measurements out of specification with the consequence of potential system mis-alignment.
Initiation date Jan 16, 2013
Report date Jun 26, 2013
Termination date Sep 30, 2016
City Andover
State MA
Country United States
FEI number 1218950