Drug recall

Beckman Coulter AU5800 Clinical Chemistry Analyzer. An automated chemistry analyzer that measures analytes…

FDA recall event 64000 · Class II · Beckman Coulter Inc.

Key facts

Event ID 64000
Recall numbers Z-0913-2013
Recalling firm Beckman Coulter Inc.
Product type Devices
Products Beckman Coulter AU5800 Clinical Chemistry Analyzer. An automated chemistry analyzer that measures analytes such as Glucose, Magnesium, and Potassium.
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Due to a software data storage limitation which only occurs under certain conditions, measurements of the Reagent Blank (RB) or Calibration (CAL) sample results are not saved and the software reverts…
Initiation date Nov 28, 2012
Report date Mar 13, 2013
Termination date Feb 24, 2014
City Brea
State CA
Country United States
FEI number 2050012

At a glance

  • 1–9 Products in this recall
  • 10–99 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date: