Drug recall
Juno DRF, X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F…
FDA recall event 63868 · Class II · Philips Healthcare Inc.
Key facts
| Event ID | 63868 |
|---|---|
| Recall numbers | Z-0702-2013 |
| Recalling firm | Philips Healthcare Inc. |
| Product type | Devices |
| Products | Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21 |
| Initiation date | Dec 13, 2012 |
| Report date | Feb 6, 2013 |
| Termination date | Jan 24, 2017 |
| City | Andover |
| State | MA |
| Country | United States |
| FEI number | 1218950 |
At a glance
- 1–9 Products in this recall
- None FDA inspections of this firm (FEI)
Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.
About FDA drug and product recalls
FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.