Drug recall

OmniGuide BP-ROBOTIC ( BeamPath Robotic), Single Use Cat. # BP-ROBOTIC The OmniGuide Fiber is indicated for…

FDA recall event 63642 · Class II · OmniGuide, Inc.

Key facts

Event ID 63642
Recall numbers Z-0453-2013
Recalling firm OmniGuide, Inc.
Product type Devices
Products OmniGuide BP-ROBOTIC ( BeamPath Robotic), Single Use Cat. # BP-ROBOTIC The OmniGuide Fiber is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Failure of glue joint, allowed a portion of the single use device to shift inside, may separate and fall into the patient
Initiation date Nov 5, 2012
Report date Dec 12, 2012
Termination date Oct 16, 2014
City Cambridge
State MA
Country United States
FEI number 3005350457

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls