Drug recall

AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585…

FDA recall event 62891 · Class II · AmeriWater Inc

Key facts

Event ID 62891
Recall numbers Z-2321-2012
Recalling firm AmeriWater Inc
Product type Devices
Products AmeriWater MRO Dialysis RO+ System, AmeriWater 1303 Stanley Avenue, Dayton, OH 45404 (800) 535-5585 www.AMERIWATER.com. The AmeriWater MRO Portable Reverse Osmosis Systems are water treatment systems…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall The firm was notified that a malfunction of the device was caused by an incorrect fuse 32 volt 15 amp fuses instead of 250 volt 15 amp fuses, that was installed in the water system. The incorrect fuse…
Initiation date Aug 9, 2012
Report date Sep 12, 2012
Termination date Apr 17, 2014
City Dayton
State OH
Country United States
FEI number 1000515504

At a glance

  • 1–9 Products in this recall
  • 1–9 FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls