Drug recall

Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal…

FDA recall event 62884 · Class II · Philips Healthcare Inc.

Key facts

Event ID 62884
Recall numbers Z-2384-2012
Recalling firm Philips Healthcare Inc.
Product type Devices
Products Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Applications
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.
Initiation date Aug 14, 2012
Report date Sep 26, 2012
Termination date Jan 17, 2017
City Andover
State MA
Country United States
FEI number 1218950

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls