Drug recall

ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and…

FDA recall event 62519 · Class II · Siemens Medical Solutions USA, Inc

Key facts

Event ID 62519
Recall numbers Z-2073-2012
Recalling firm Siemens Medical Solutions USA, Inc
Product type Devices
Products ARTISTE or ONCOR Digital Linear Accelerators with syngo RT Therapist version 4.2/MOSAIQ OIS Systems and Multiple-X, Manufactured by SIEMENS AG. The ARTISTE and ONCOR family of Linear Accelerators are…
Classification Class II
Status Terminated
Voluntary or mandated Firm Initiated
Reason for recall Potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare…
Initiation date Jun 25, 2012
Report date Aug 1, 2012
Termination date Oct 18, 2012
City Concord
State CA
Country United States
FEI number 2910081

At a glance

  • 1–9 Products in this recall
  • None FDA inspections of this firm (FEI)

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About FDA drug and product recalls

FDA enforcement reports list recalls of drugs, biologics, foods and other FDA-regulated products, classified by the potential health hazard.

Source: U.S. Food and Drug Administration (FDA), Enforcement Reports · Source date:

Browse all FDA drug and product recalls