Drug recall
Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor…
FDA recall event 62437 · Class II · Spacelabs Healthcare, Llc
Key facts
| Event ID | 62437 |
|---|---|
| Recall numbers | Z-2019-2012 |
| Recalling firm | Spacelabs Healthcare, Llc |
| Product type | Devices |
| Products | Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or alarms. The Spacelabs Healthcare Xprezzon Bedside Monitor is… |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Firm Initiated |
| Reason for recall | There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display… |
| Initiation date | Jun 12, 2012 |
| Report date | Jul 25, 2012 |
| Termination date | Nov 28, 2012 |
| City | Issaquah |
| State | WA |
| Country | United States |