CFR section

21 CFR 820.30(j): Design controls

PART 820-QUALITY SYSTEM REGULATION · Subpart C-Design Controls

Key facts

Citation 21 CFR 820.30(j)
Section Design controls
Part PART 820-QUALITY SYSTEM REGULATION
Subpart Subpart C-Design Controls
Subchapter 21 CFR H Medical Devices
Product types cited Devices, Drugs, General, Other_Regulations
Regulation text (excerpt) (j) Design history file. Each manufacturer shall establish and maintain a DHF for each type of device. The DHF shall contain or reference the records necessary to demonstrate that the design was developed in accordance with the approved design plan and the requirements of this part.

At a glance

  • 100–999 Inspections citing this section
  • 100–999 Warning letters citing this section

Counts are shown as ranges. Exact figures, trends and every underlying record are available in Orca1.

About CFR sections cited by the FDA

Title 21 of the Code of Federal Regulations contains the FDA's regulations, including current good manufacturing practice and quality system requirements that inspections cite.

Source: U.S. Code of Federal Regulations (eCFR); citation statistics from FDA inspection and warning-letter data

Browse all CFR sections cited by the FDA